Guidance and Manufacturer's Declaration: Electromagnetic Immunity
1288 HD Camera is intended for use in the electromagnetic environment specified below.
The customer or the user of 1288HD Camera should ensure that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD) IEC61000-4-2
Electrical fast transient/
burst IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
NOTE: Ut is the AC mains voltage prior to application of the test level.
42
IEC 60601 Test Level
±6kV contact
±8kV air
±2kV for power supply
lines
±1kV for input/output
lines
±1kV differential mode
±2kV common mode
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
3 A/m
Compliance Level
±2,4,6kV contact
±2,4,8kV air
±2kV line to ground
±1kV line to line
±0.5, 1kV differential
mode
±0.5, 1, 2kV common
mode
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
N/A
Electromagnetic
Environment:
Guidance
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of 1288
HD Camera requires
continued operation
during power mains
interruptions, it is
recommended that
1288 HD Camera
be powered from an
uninterruptible power
supply or a battery.
Power-frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.