Sterrad®
1. Clean and prepare the camera head and cable as recommended in this user guide. Ensure the
soaking cap is installed.
2. Double wrap the camera head and cable prior to sterilization.
3. Sterilize the camera head and cable following the instructions of the manufacturer, using the
STERRAD® 100S, 200, NX™, or 100NX™ Sterilization System. Select the standard cycle.
Warning: Not all sterilization trays are compatible with STERRAD® systems. Using an incompatible
tray may result in incomplete device sterilization. Consult the instructions that came with your
sterilization tray to determine which sterilization method is compatible with your tray and
devices. If a compatible tray is not available, the devices can be double-wrapped prior to using the
STERRAD® system.
4. Allow the camera head, cable, coupler, and scope to completely dry before reassembly. Any
moisture on the threads will cause the camera and coupler windows to fog during use.
Storage
Never store the device in a non-ventilated, humid environment such as a carrying case. This may
present an infection control risk.
Using Sterile Drapes
Using sterile drapes will ensure maximum longevity of your 1288 HD Camera Head. For best results,
follow the instructions provided by the drape manufacturer.
37