User Maintenance

Replacing the Fuses

To avoid the risk of fire, use only fuses of the value specified on the fuse label
located on the rear panel of the transmitter.
1. Unplug the power cord from the wall outlet and remove the cord from the transmitter console.
2. Unlatch the fuse holder above the AC inlet and remove it. (You may need to press the tab on the
fuse holder with a slender screwdriver to release the latch.)
3. Replace the fuse with the same value and rating.
4. Reinstall the fuse holder until the tab snaps in place.

Periodic Maintenance Schedule

To ensure safe operation of the Model 1288 HD Video Camera you should
periodically perform the following procedure:
Every 12 months, check the earth leakage current to <500µA (<300µA in U.S.A.), ground protective
earth impedance to <0.1 ohms, and power consumption less than or equal to rated power. Use a true
RMS digital multimeter and safety analyzer to perform this test.
Note: Refer calibration and operating difficulties not detailed in this manual to your Stryker
Endoscopy sales representative.
Disposal
This product contains electrical waste or electronic equipment. It must not be
disposed of as unsorted municipal waste and must be collected separately in
accordance with applicable national or institutional related policies relating to
obsolete electronic equipment.
The 1288 HD must be disposed of according to local laws and hospital practices.