Limitations on Reprocessing
•
Do not cross-sterilize the device. Using multiple sterilization methods may significantly
reduce the performance of the device.
•
Do not leave the device in solutions longer than necessary. This may accelerate normal
product aging.
•
Proper processing has a minimal effect on this device. End of life is normally determined
by wear and damage due to use.
•
Damage incurred by improper processing will not be covered by the warranty.
Instructions
Point of Use
•
Wipe excess soil from the device using disposable paper towels.
•
If an automated reprocessing method will be used, rinse any channels in the device with
50mL of sterile distilled water immediately after use.
Containment and Transportation
•
Reprocess the device as soon as reasonably practical following use
•
Transport the device in a tray to avoid damage.
A 30 minute wait time was used during cleaning validation.
1
Preparation for Cleaning
1. Disassemble the coupler from the scope and camera head.
2. Prepare an enzymatic detergent according to the manufacturer's recommendations (one ounce
per gallon of tap water at 35 - 40°C)
3. Wipe the entire device with the detergent, using a clean cloth.
4. Immerse the device in the detergent. Using a syringe, inject any inside regions of the device
with 50mL of the detergent to ensure all parts of the device are reached.
5. Soak the device in the detergent for a minimum of 15 minutes.
32
.
2
.
1