Ethylene Oxide (EO)
Double wrap camera head and cable prior to sterilization.
•
Preconditioning parameters
Temperature
Chamber Humidity
Vacuum Set Points
Time
Exposure
Concentration (100% EO)
Temperature
Time
Chamber Humidity
Aeration parameters
Aeration Time
Temperature
Steris® System 1 / System 1E
Note: Steris® System 1™ is not intended for use in the United States.
1. Clean and prepare the camera head and cable as recommended in this user guide. Ensure the
soaking cap is installed.
2. Sterilize the camera head and cable using Steris® System 1™ with Steris® 20 Sterilant, or System
1E™ with S40™ Sterilant. Follow the manufacturer's instructions.
3. Allow the camera head, cable, coupler, and scope to completely dry before reassembly. Any
moisture on the threads will cause the camera and coupler windows to fog during use.
Steris/Amsco V-PRO 1
1. Clean and prepare the camera head and cable as recommended in this user guide. Ensure the
soaking cap is installed.
2. If using a sterilization tray (optional), follow any additional instructions provided with the
tray. Use only trays that are approved for sterilization with Steris/Amsco® V-PRO™ 1.
3. Double-wrap the camera head and cable (or tray, if being used) prior to sterilization.
4. Sterilize the camera head and cable using the V-PRO™ 1 Plus Sterilizer (Non-Lumen or
Standard cycle) or the V-PRO™ 1 Sterilizer (Standard cycle).
5. Allow the camera head, cable, coupler, and scope to completely dry before reassembly. Any
moisture on the threads will cause the camera and coupler windows to fog during use.
36
55°C (131°F)
70% RH
1.3 psia
30 minutes
725 mg/L
55 ± 2°C (131 ± 5°F)
1 hour
70% RH (50–80%) ± 5%
12 hours
35–54°C (95–129°F)