Problem
Optics are dirty
Blurry picture
Note:
If this troubleshooting guide does not resolve the problem, call Stryker Technical Support at
1-877-478-7953 (inside the U.S.) or refer to the "Warranty and Return Policy" section of this manual.
30
Possible Solution
•
Rotate the scope. If dust particles in the picture rotate, the dust is
located on the scope itself. Follow the manufacturer's instructions for
cleaning the eyepiece and negative lens.
•
If particles in the picture do not move when you rotate the scope, the
particles are located on the coupler or camera. Remove the scope and
clean the window on the front of the coupler with a dry or alcohol-
tipped cotton swab.
•
If dust particles lie between the coupler and camera, remove the coupler
and clean the coupler and camera windows.
•
Ensure all components are completely dry before reassembling them, or
fogging may result.
•
Ensure the coupler or C-Mount scope is in focus.
•
Increase the enhancement.
•
Ensure the specialty switch is not set to FLEXI-SCOPE unless you are
using a flexible scope.