Operating The Device; Supplies Required; Entering Case Data - Medtronic autoLog IQ Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Size: 7 x 9 in (178 x 229 mm)
2. Press FCT to allow for the manual transfer of volumes or to remove air from the holding bag. Choose from the 3 options shown in
Figure 21.
Figure 21. FCT capabilities screen
1 FCT 1: Empty the bowl to the reservoir. Allows for the manual transfer of blood to the reservoir.
2 FCT 2: Empty the bowl to the holding bag. Allows for the manual transfer of blood to the holding bag.
3 FCT 3: Invert the holding bag. Allows for the manual removal of air from the holding bag.
3. Press and hold
Warning: Failure to press and hold
4. Press
at any time to return to the Main screen.

13 Operating the device

13.1 Supplies required

Warning: Only use Medtronic wash kits and blood collection reservoirs with the device.
Note: The Medtronic disposables are sterilized using ethylene oxide.
The following supplies are required:
• Medtronic wash kit that includes the following:
– Centrifuge bowl (135 mL capacity) with tubing
– Holding bag (1000 mL capacity)
– Waste bag (10 L capacity)
• Medtronic blood collection reservoir (4 L capacity)
• Adequate quantity of 0.9% isotonic saline for washing
• Adequate quantity of anticoagulant solution
• Blood transfer bags
• Suction/anticoagulant line

13.2 Entering case data

1. Press
.
Figure 22. Case data entry screen
2. Press the text field corresponding to the type of data being entered.
3. To manually enter data, use the keyboard to enter the data into the selected text field.
M967226A006 1B Medtronic Confidential
to perform the selected Function.
when inverting the holding bag can result in air not being removed from the bag.
25
502431-132
English
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018

Publicidad

Tabla de contenido
loading

Tabla de contenido