Device Disposal And Ownership Transfer; Deleting Case Data; End Of Life Disposition; Specifications - Medtronic autoLog IQ Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Size: 7 x 9 in (178 x 229 mm)

20 Device disposal and ownership transfer

20.1 Deleting case data

Delete all case data before disposing or transferring ownership of the device (Section 14.5).

20.2 End of life disposition

Contact your Medtronic service representative before disposing of the device.

21 Specifications

21.1 Technical specifications

Electrical classification
Power
Voltage
Frequency
Phase
Current
Fuses
Power cord
Speed, flow rate, and pressure
Centrifuge
Pump
Vacuum
Weight sensor
Dimensions
Width
Height (without IV pole)
Depth
Weight (device including IV pole)
IP rating
Temperature limit
Operational
Storage (clinic)
Storage (warehouse)
Transit
Humidity range
Operational
Storage (clinic)
Storage (warehouse)
Transit
Pressure range
Operational
Storage (clinic)
Storage (warehouse)
Transit

21.2 Electromagnetic compatibility declaration

• IEC 60601-1-2, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance –
Collateral Standard: Electromagnetic disturbances – Requirements and tests.
• Class A, Group 1
Warning: The emissions characteristics of this device make it suitable for use in industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential environment (for which CISPR 11 class B is normally required) this device might not offer adequate
protection to radio frequency communication services. The user might need to take mitigation measures, such as relocating or
reorienting the device.
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should
be used no closer than 30 cm (12 in) to any part of the device, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this device could result.
Warning: Do not stack or attach additional instrumentation to the device, in order to minimize the potential for RF interference.
Warning: Power cord and USB cable length must not exceed 3 m (118 in).
M967226A006 1B Medtronic Confidential
Class I, Type BF (suction/anticoagulant line), Ordinary, Contin-
uous operation
100 V~ to 240 V~
50 Hz / 60 Hz
Single
10 VA to 425 VA
7 A / 250 V slow blow, 3AG, 200 A breaking capacity (Littelfuse
0313007.MXP or equivalent)
3 prong hospital grade connector (varies by geography); length
must not exceed 3 m (118 in)
0 rpm to 10 000 rpm (±5%)
0 mL/min to 1 000 mL/min (±7%)
−10 mmHg to −370 mmHg ±( |[5%]| + 8 mmHg)
Self-start: 800 mL ±400 mL
69 cm (27 in)
80.5 cm (31.7 in)
42.5 cm (16.7 in)
50 kg (110 lb)
IPX1
15°C to 30°C (59°F to 86°F)
15°C to 30°C (59°F to 86°F)
15°C to 30°C (59°F to 86°F)
-35°C to 60°C (-31°F to 140°F)
25% to 70% noncondensing
25% to 70%
10% to 90%
10% to 90%
80 kPa to 101 kPa (11.6 psi to 14.6 psi)
80 kPa to 101 kPa (11.6 psi to 14.6 psi)
80 kPa to 101 kPa (11.6 psi to 14.6 psi)
59.5 kPa to 101 kPa (8.6 psi to 14.6 psi)
51
502431-132
English
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018

Publicidad

Tabla de contenido
loading