Size: 7 x 9 in (178 x 229 mm)
Component/area
Storage tray
Internal storage compart-
ment
Touchscreen
Barcode scanner
17 Special blood processing conditions
Note: The following steps should be followed in conjunction with normal setup steps in Section 13.3.
17.1 Emergency setup using multiple suction/anticoagulant lines on 1 reservoir
The setup for this configuration is similar to that used for a normal case.
M967226A006 1B Medtronic Confidential
Cleaning
1. Any time the storage tray becomes dirty (for example, blood spills), it should be removed and
cleaned according to approved hospital protocols with a 10% bleach solution or equivalent
disinfectant solution.
2. After cleaning, wipe the tray with a cloth and water to remove any cleaning solution residue. Do
not use alcohol.
3. Allow the storage tray to dry before reinstalling.
1. Whenever the internal storage compartment becomes dirty (for example, blood spills), it
should be removed and cleaned.
2. To remove the storage compartment, open it.
3. Retract the spring plunger in the upper left corner on the inside of the compartment, and open
the compartment completely.
4. Lift the compartment off of the 2 pins in the device front panel.
5. Clean the compartment and the area inside the device from which the compartment was
removed according to approved hospital protocols using a 10% bleach solution or equivalent
disinfectant solution.
6. After cleaning, wipe with a cloth and water to remove any cleaning solution residue. Do not use
alcohol.
7. Allow the storage compartment and device to dry before reinstalling.
8. To reinstall the compartment, align the sockets on the bottom of compartment with the 2 pivot
pins in the device front panel.
9. While closing the compartment, retract the spring plunger in the upper left corner of the com-
partment. Release the spring plunger once it has passed the compartment stop plate.
Clean the touchscreen with a soft, lint-free cloth.
Warning: Do not use bleach to clean the touchscreen, as this could discolor the screen and result
in failure.
For barcode scanner cleaning instructions, refer to the literature included with the scanner.
37
502431-132
English
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018