Figure 27. Wash volume screen
13.5.2 Pausing and stopping blood processing
If necessary to pause a Fill, Wash, or Empty phase, press . During the Fill and Wash phases, the centrifuge continues to spin,
and the pump pauses for 60 s. Press
selected within 60 s.
If necessary to stop a Fill, Wash, or Empty phase, press . This automatically returns the bowl contents to the reservoir.
Note: Function capabilities are only enabled when the device is in Stop mode.
13.5.3 Blood transfer
1. If necessary, empty the holding bag line by pressing FCT 2. The following prompt appears: FCT 2: Empty the bowl to the
holding bag. Press and hold
2. Transfer blood from the holding bag to a transfer bag, transfer blood in the holding bag, or transfer blood by direct patient
connection.
Warning: If a direct patient connection to the processing device is required, take additional measures to detect and prevent air
embolism (refer to the current version of AABB Standards for Perioperative Autologous Blood Collection and Administration).
Note: If using a transfer bag, remove all air from the transfer bag line by squeezing the transfer bag, before transferring it to the
anesthesiologist.
Note: If using the holding bag to transfer blood, use FCT 3 to remove air from the bag, before transferring it to the
anesthesiologist.
3. Place a new holding bag on the IV pole and connect it to the device if applicable.
4. Press
to continue.
13.6 Resuming a case
If the device power has been turned off during a case, resume the case by restarting the device within 6 hours. To resume a case,
perform the following steps:
1. Press
to turn on the device. The power LED illuminates, the alert light flashes, and the speaker emits 2 beeps.
2. The following message appears: Resume previous case? If Yes, bowl contents will be pumped to reservoir. Make sure
the reservoir is connected. To resume the previous case, select Yes.
Note: The device defaults to a new case if the device power is turned on 6 or more hours after shutdown or power loss.
14 Managing case data
A password must be set in order to store and access case data.
The device stores data for up to 100 cases. A notice appears when 90 cases have been stored, to inform the user that the storage
is nearly full. If additional case data is stored after the limit of 100 cases has been met, 1 old case will be deleted for each new case
added.
Refer to institutional policies regarding the handling of protected health information and data security.
14.1 Viewing case data
1. To view current case data, press Data.
Note: To view all case data, press Login, enter the password, and press
Note: After the password has been entered, cases appear chronologically, with the most recent case on top. A scroll bar
appears if more than 6 cases are displayed.
2. To filter the data, see Section 14.2.
3. Select the desired cases by pressing anywhere in the applicable rows.
Note: Press Select all to see data for all stored cases. Press Select none to undo this.
M967226A006 1B Medtronic Confidential
to resume. The device times-out and returns the bowl contents to the reservoir if
to run the pump. Press
when complete. (Section 12.7).
31
. If a password has not been set, see Section 12.4.