17. If necessary, adjust the vacuum setting using the Vacuum drop-down menu on the Main screen. The vacuum setting can be
increased or decreased manually using
Section 12.1 to set the Default and Max vacuum settings.
Warning: Setting the vacuum at higher than desired settings, or unintentionally selecting Max vacuum, could result in
hemolysis and tissue damage.
Note: The factory default vacuum setting is −120 mmHg.
Note: After selecting Max vacuum, the message Confirm max vacuum? appears. Select Yes to confirm or No to decline.
Note: Selecting Max vacuum causes a notification to appear every 10 min. If the notification is postponed, the notification will
reappear at the interval selected in the Reset alert field.
Figure 25. Vacuum settings screen
18. Before blood collection, open the roller clamp and prime the collection reservoir with a minimum of 200 mL of anticoagulant
solution.
19. Reduce the anticoagulant flow to a ratio of approximately 15 mL of solution to 100 mL of blood.
20. Monitor the collection reservoir periodically for appropriate anticoagulation.
21. Prime the collection reservoir with anticoagulant solution each time the collection reservoir is emptied.

13.4 Setting up the wash kit

See Table 1 for prompts associated with the setup of the disposables.
1. Attach the saline wash solution containers to a lower IV pole hook.
2. Rotate the blue wash kit cradle arm as shown in Figure 26.
3. Place the wash kit in the cradle arm and remove the wash kit cover.
Figure 26. Wash kit cradle arm
4. To enter the patient and operator IDs, and the disposable lot, heparin lot, and ACD-A lot data into the device, see Section 13.2.
5. Remove the waste bag from the wash kit.
6. Confirm that the drain valve is closed on the waste bag.
Warning: Failure to close the drain valve can lead to blood leakage.
7. Install the waste bag onto the waste bag pins on the right-facing side of the device. Ensure that the waste bag volume label faces
away from the device.
8. Remove the holding bag from the wash kit and hang it on an upper IV pole hook. Close the clamps on the 2 red-capped outlet
lines that each include a membrane port.
9. Place the bowl in the centrifuge chamber with the yellow-capped waste line outlet tubing facing the waste bag. Refer to the
markings on the top panel.
10. Align the centrifuge markings with the vertical assemblies on the bowl.
11. To lock the bowl, simultaneously press down on the bowl and turn it clockwise until a click is heard.
12. Insert the yellow-capped connector into the yellow-capped inlet port on the waste bag.
English
M967226A006 1B Medtronic Confidential
and , or it can be set to the preprogrammed Default or Max vacuum settings. See
28