Size: 7 x 9 in (178 x 229 mm)
14.5 Deleting case data
1. To delete cases, select desired cases, and then press Delete on the case list screen.
2. Press Yes to accept the deletion. Press No to reject the deletion.
Figure 34. Delete selected cases screen
15 Ending a case
1. If necessary, turn off the vacuum by pressing
2. Document the case results.
3. Turn off the device by pressing
4. Cover all exposed ports and connectors of the disposables.
5. Close all tubing clamps.
6. Unlock and open the manifold cover.
7. Pull the pump lever and remove the tubing from the pump.
8. Remove the reservoir, centrifuge bowl, tubing harness, waste bag, vacuum line, and all solution containers from the device.
Warning: Only clamp the waste line after the disposables have been removed.
9. Follow hospital procedures when disposing of biohazardous waste.
10. Clean the device in accordance with hospital protocols and as described in Chapter 16.
16 Cleaning the device
Warning: Use appropriate personal protective equipment such as eye protection, mask, and gloves when cleaning the device and
discarding used fluids. Failure to do so could result in exposure to bloodborne pathogens, blood, and cleaning fluids.
Warning: If it is suspected that fluid has entered the device, it should be immediately examined by a trained technician. Fluid ingress
could result in electric shock to the user.
Warning: Clean and maintain the device in accordance with this manual and hospital procedures.
Component/area
External
Level sensor
Air detector and valve
assembly
M967226A006 1B Medtronic Confidential
.
for 1.5 s. Unplug the power cord from the supply mains.
Cleaning
1. Any time the external portions of the device become dirty (for example, blood spills), they
should be cleaned according to approved hospital protocols with a 10% bleach solution or
equivalent disinfectant solution.
2. After cleaning, wipe the device with a cloth and water to remove any disinfectant solution
residue. Do not use alcohol.
3. Allow the device to dry.
If blood or other solutions are spilled in the centrifuge chamber, or dust builds up on the level sensor
during use, the sensor can become inoperative.
1. To clean the level sensor, use a foam-tipped swab dipped in a mild detergent and water.
Caution: Do not use liquids that may leave a film, such as alcohol. Do not use an abrasive
cloth.
Note: To clean blood spills, use a 10% bleach solution or equivalent disinfectant solution.
2. Carefully wipe across the sensor to remove any blood, dust, or debris.
3. Carefully wipe across the sensor with another foam-tipped swab to remove any remaining
moisture.
1. To clean the air detector or valve assembly, use a foam-tipped swab dipped in a mild detergent
and water.
Caution: Do not use liquids that may leave a film, such as alcohol. Do not use an abrasive
cloth.
35
502431-132
English
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018