Setting Up The Device For Blood Collection - Medtronic autoLog IQ Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Size: 7 x 9 in (178 x 229 mm)
Note: If using multiple disposables for a case, record the additional data in the Notes text field using the barcode scanner or
keyboard.
4. Press
on the keyboard to enter the data in the device.
5. To scan data into the device using the optional barcode scanner (Chapter 10), press
6. Press the text field corresponding to the type of data being scanned.
7. To use the scanner, aim it at a barcode. Ensure that the red scanner line fully covers the barcode, and then press the trigger.
Note: The scanner can read the following barcode formats, at a minimum: aztec, code 128, code 39, data matrix ECC 200, and
PDF 417.
Note: The scanner cannot read custom barcode formats.
8. After a successful scan, the scanner beeps, and the green LED illuminates. If unsuccessful, the red LED illuminates.
9. Press
on the screen to record the scanned data in the device.
Note: For additional barcode scanner information, refer to the literature included with the scanner.
10. If desired, press Main to return to the Main screen.

13.3 Setting up the device for blood collection

Figure 23. Device components
1 Anticoagulant solution
2 Saline bag (1 or more)
3 Holding bag
4 Manifold (3-way tubing assembly)
5 Line to holding bag
6 Line from saline bag
7 Line from blood collection reservoir
Warning: Failure to adhere to the following loading conditions when using the device may lead to tipping:
• When using the bottom IV pole hooks, do not attach more than 6 L of fluid, and do not attach any bag higher than 48.3 cm (19 in)
above the device.
• When using the top IV pole hooks, do not attach more than 3 L of fluid, and do not attach any bag higher than 86.4 cm (34 in)
above the device.
• 2 collection reservoirs may be filled with no more than 4 L of fluid in each.
• 1 waste bag may be filled with no more than 7 L of fluid.
English
M967226A006 1B Medtronic Confidential
on the screen.
8 Step-down connector
9 Suction/anticoagulant line to patient
10 Blood collection reservoir
11 Vacuum line to the vacuum overflow trap of the device or to
external regulated vacuum source
12 Waste bag
26
502431-132
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018

Publicidad

Tabla de contenido
loading

Tabla de contenido