en
®
7.4.3
OrthoPilot
4 wireless foot switch
Fig. 3
Information, such as article number, serial number and firmware status is
displayed via the "Camera" button.
Commissioning (only necessary when connecting a new foot switch)
®
►
Switch on the OrthoPilot
bottom to the "ON" position "I".
Note
The on/off switch is provided solely as a transport safeguard and can remain
in the ON "I" position after initial startup.
►
Select the "Camera" button.
►
Select the search button in the dialog window.
®
A list of the OrthoPilot
4 cameras found appears.
Note
If multiple devices are displayed, the foot control can be identified by the
serial number. The serial number is located on the type plate on the back of
the camera.
®
►
Select the OrthoPilot
4 camera with the right serial number.
Successful connection will be displayed.
®
The OrthoPilot
4 wireless camera is now ready to use.
Note
The range of the radio connection between the OrthoPilot
®
era and the OrthoPilot
4 panel PC unit is max. 5 m.
10
4 wireless camera: turn the ON/OFF on the
®
4 wireless cam-
7.5
Settings
System settings can be configured in this view.
a
b
c
d
e
f
g
h
Fig. 4
The window is divided into the following area:
Designation
a
Hospital
b
Language
c
Time Zone
d
Layout +
Personalization
e
Layout Appearance Select background color (black or white).
f
Sound
g
Data Privacy
h
Info
Description
Enter hospital information.
Select language (English, German, Span-
ish)
Set time zone and summer or winter time.
Edit Mode: Changes to the tile size and
tile name can be made on the home
screen, see Chapter 7.3.
Note: Applications/application processes
cannot be started in the edit mode.
User Mode: Standard mode; Changes
cannot be made.
Manage volume settings:
■
Sound on/off
■
Playback volume
Information on Aesculap data protection
Display of patents and licenses