9. INSTALLATION AND COMMISSIONING
For first use, check that:
• Packaging is intact and has protected the device during transportation.
• Check that all parts included in the packing list are present
• General functionality of the device as described in the instructions for use.
• The medical vehicle is equipped with suitable surfaces to ensure that the requirements of EN 1789 are met and that installation does not impair access to other devices
Do not under any circumstances modify any part of the hook as this could cause damage to the patient and/or operators.
Failure to take the above measure will preclude safe use of the device, resulting in risk of damage to the patient, operators and the device itself.
It is advisable to practise in non-operational conditions before regular commissioning. For subsequent use, perform the operations specified in paragraph 12.
If the above conditions are met, the device may be considered ready for use; otherwise, you must immediately remove the device from service and contact the Manufacturer.
Do not alter or modify the device arbitrarily, as doing so could result in unpredictable operation and damage to the patient or rescuers and will void the warranty and
release the Manufacturer from all liability.
10. OPERATING CHARACTERISTICS
IT
11. PROPER USE
11.1 EMERGENCY VEHICLE REQUIREMENTS
EN
The Fastening system must be properly installed when positioned inside the ambulance in order to limit the risks to the vehicle occupants that could be generated by violent
acceleration or deceleration of the vehicle. For this purpose, the emergency vehicle must be equipped with suitable surfaces capable of firmly holding the chair and fastening
system assembly.
Improper installation of the fastening system can result in serious injury to the vehicle occupants.
DE
11.2 INSTALLATION
For correct installation of the Fastening system, the installation surfaces must be perpendicular to each other and level at all fixing and chair support points. Failure to comply
with these conditions may result in altered geometries that do not conform to the design specifications.
The fastener is provided with 4 holes to be used for wall installation.
FR
Before drilling the holes necessary for installation, ensure that no part, element
or device within the vehicle can be compromised or damaged during this step.
Take special care to ensure the integrity and serviceability of electrical circuits and
components, structural parts and pneumatic circuits.
ES
To this end, the following pictures show the drilling dimensions required for
installation.
PT
DIMENSIONS
A
102.5 mm
PL
B
200 mm
C
697.5 mm
D
450 mm
BG
Drill ø6mm holes using the fastener as a drilling template.
A counter plate (not supplied) must be placed on the back of the installation surface
RO
to ensure secure installation of the device.
Complete installation using TCEI M6 screws of the appropriate length for the wall
thickness, washers and self-locking nuts on the back side.
14
ELEMENT
DESCRIPTION
1
Lower stop elements
2
Spacers (only on 4Bell+ Max)
3
Main body
4
Quick-release buckle
FUNCTION
Needed to hold the chair in position by holding its
lower part
Spacers used to stabilise the 4Bell+ chair
Structural part of the fixing, required for wall fastening
Allows the belt to be fastened around the chair
D
C
A
B