If the acquisition is interrupted before its completion, the entire portion of analysed tissue will however be shown on
the PC together with an error message.
If the image is exhaustive, even if partial, the examination does not need to be repeated, thus avoiding subjecting the
patient to a further dose of radiation.
In order to comply with the operator safety regulations on ionizing radiation (for clarifications, see paragraph 1.8.7
"Protection against ionizing radiation"), the operator must stand at least 2 metres away from the X-ray device during
emission.

4.3. PERFORM A SIMULATION (DUMMY RUN)

The unit moves when the following operations are performed. The operator must:
- Closely monitor the patient and immediately let go of the button to stop movement if there is a risk of the
device hitting the patient.
-
Prevent the patient from assuming incorrect positions (putting hands or other body parts in inappropriate
areas) or moving out of the examination area.
To perform a simulated cycle, a simulation of the projection mode or program selected, but without emission of radiation
(dummy run), hold down the button
This button must be held down for the entire cycle, otherwise the movement will stop.
This simulation can be useful for particularly emotional patients, especially children, showing them what the
examination consists of without having to leave them alone in the room.
EN
(CONFIRM).
OPERATOR'S MANUAL
23