10. ERROR MESSAGES

Code
WARNING
W 0.1
W 0.2
W 0.3
W 0.5
W 0.7
W 0.8
W 0.9
W 0.10
W 0.13
W 0.14
W 0.15
Code
LOGIC PANEL ERRORS
E X.X
EN
Message
X-ray
button
released
exposure.
X-ray button not released at the end
of the examination.
Emergency button pressed.
POWER electronic card can not
communicate with Logic card.
Active interlock. The door has been
opened with the examination in
progress.
Demo
mode:
examination
loaded.
The unit is not set up.
The unit is not calibrated.
Button pressed on the keyboard
upon turning on or wake-up after
standby.
Unit temperature too high.
X-ray generator temperature too
high.
Message
Internal error
OPERATOR'S MANUAL
Description/Solution
during
Hold down the X-ray emission button
until the end of the procedure.
1) X-ray button pressed for too
long, after the end of the
exposure (>15s).
2) Reset command output when X-
ray button was pressed.
3) The device remained in standby
mode when X-ray button was
pressed.
Release the emergency button.
Turn off the device, wait for 30
seconds and turn it on again. If the
error persists, check the error code
documentation.
Close the door and, if necessary,
repeat the examination.
not
Demo started without loading the
examination: check connection with
the PC. Repeat the operation.
If the error persists, check the error
code documentation.
If the error persists, check the error
code documentation.
Check the keyboard. If the error
persists,
check
documentation.
Temperature of converter card too
high, it is not possible to start a new
examination. Wait for cooling.
Temperature of X-ray generator too
high, it is not possible to start a new
examination. Wait for cooling.
Description/Solution
If the error persists, check the error
code documentation.
the
error
code
61