The control power supply fuse always burns out.
The engine pump won't stop.
The platform does not open or opens slowly, the engine
pump is running.
The platform does not open or opens slowly, the engine
pump is running.
The platform goes down too quickly or too slowly (maxi-
mum speed 5.9 ips).
The platform goes down on its own remaining horizontal
(i.e. 0.8-1.2 in in 5 minutes).
The platform does not go down.
The platform does not go up or down with the pedal
buttons .
The platform tilts the tip by itself (i.e. 1.2-2 in in 5
minutes).
The platform opens too quickly.
The platform does not close completely.
The control power supply fuse always burns out.
The engine pump won't stop.
The platform does not open or opens slowly, the engine
pump is running.
The platform does not open or opens slowly, the engine
pump is running.
The platform goes down too quickly or too slowly (maxi-
mum speed 5.9 ips).
The platform goes down on its own remaining horizontal
(i.e. 0.8-1.2 in in 5 minutes).
The platform does not go down.
The platform does not go up or down with the pedal
buttons .
The platform tilts the tip by itself (i.e. 1.2-2 in in 5
minutes).
The platform opens too quickly.
The platform does not close completely.
En
Short circuit or excessive electricity absorption.
The power relay remains inserted.
The EVSR cylinder valves of the tilting device are not
responding.
VFR flow adjusting valve faulty.
Platform is frozen against the body frame.
Platform pivots hard.
The platform scrapes against the frame.
VFS flow adjusting valve is dirty or faulty.
Lack of tightness in the cylinder lifting seals.
Cables or valves EVSS of the lifting cylinders or valve EVBS
are defective.
Foot control switch is defective .
Interruption of platform cable.
Command selector switch not activated.
Lack of tightness in the cylinder tilting seals.
VFR flow adjusting valve is dirty or faulty.
Tilt cylinder rod extensions poorly regulated.
En
Short circuit or excessive electricity absorption.
The power relay remains inserted.
The EVSR cylinder valves of the tilting device are not
responding.
VFR flow adjusting valve faulty.
Platform is frozen against the body frame.
Platform pivots hard.
The platform scrapes against the frame.
VFS flow adjusting valve is dirty or faulty.
Lack of tightness in the cylinder lifting seals.
Cables or valves EVSS of the lifting cylinders or valve EVBS
are defective.
Foot control switch is defective .
Interruption of platform cable.
Command selector switch not activated.
Lack of tightness in the cylinder tilting seals.
VFR flow adjusting valve is dirty or faulty.
Tilt cylinder rod extensions poorly regulated.
Detect and eliminate the short circuit.
Switch off the valve coils to trace the faulty one.
Turn off the main switch.
If necessary, replace the power relay.
Check the valve coils, cables, the signal output on the
printed circuit command and on the valve body.
Replace valve.
Defrosting, treat the seal to prevent freezing.
Remove and clean the platform pivots.
Turn on briefly the Lowering, then the Opening.
Replace valve.
Replace the cylinder seals.
Replace the cables and/or valve.
Replace the switch.
Repair or replace the cable.
Turn on the selector.
Replace the cylinder seals.
Replace valve.
Adjust properly the extension of the cylinders.
Detect and eliminate the short circuit.
Switch off the valve coils to trace the faulty one.
Turn off the main switch.
If necessary, replace the power relay.
Check the valve coils, cables, the signal output on the
printed circuit command and on the valve body.
Replace valve.
Defrosting, treat the seal to prevent freezing.
Remove and clean the platform pivots.
Turn on briefly the Lowering, then the Opening.
Replace valve.
Replace the cylinder seals.
Replace the cables and/or valve.
Replace the switch.
Repair or replace the cable.
Turn on the selector.
Replace the cylinder seals.
Replace valve.
Adjust properly the extension of the cylinders.
183
183