ii. Insert the tip of the catheter into the urethral orifi ce, and then place the front end of the Advancer against the external urethral orifi ce to locate the
catheter. Adjust the angle of the catheter to ensure that the catheter is coaxial with the urethra, which can reduce the friction with the urethra in the
process of pushing the catheter (See Fig.1).
The tip of the catheter is aimed at the
urethral orifice.
Urethra
iii. Push the catheter into the bladder at constant speed and slowly with the Advancer. Keep pushing 3-5 cm after you see the urine (See Fig.2). If the
Advancer is pushed to the maximum position and still has no urine, remove the Advancer and continue to push the catheter 3 to 5 cm until the urine
is seen.
iv. The catheter is gently drawn outward after fi lling the Foley balloon with the syringe. The resistance can prove that the balloon has been fi xed at the
neck of the bladder, and fi nally the Advancer is removed. (See Fig.3)
v. Injection water or physiological saline is injected into the balloon from the valve, and then gently pull the catheter outwards. Encountering resistance
can prove that the balloon has been fi xed to the neck of the bladder. (See Fig.4)
FIG.3
vi. The following operations are consistent with the standard type Foley catheter.
Warning:
1.
This product is only for single use. Reusing of used product is strictly prohibited. Used product should be properly disposed as medical waste or
biohazardous waste according to local hospital protocol.
2.
This product is sterilized with ethylene oxide and period of validity is 2 years.
3.
Handle the product gently with care and do not pull it hard when used.
4.
The water injection of the balloon should not exceed the maximum volume of the balloon.
5.
This product should be operated by professional medical personnel.
6.
Do not use lubricants with oil matrix.
7.
Check the condition of package and confi rm whether the product is in the validity period before use. Do not use if the product has expired or the package
is damaged.
8.
If the female type Foley catheter cannot be pushed normally during the process of using Advancer, disassemble the Advancer. Replace the Advancer
with tweezers according to the usage method of standard type catheter.
9.
0.9% sterilized physiological saline can be used in the irrigation lumen of the catheter to irrigate the bladder.
Indwelling Time:
This product should not be retained in the body for more than 14 days.

Storage Environment:

The product should be stored in a ventilated, dry, non-corrosive room. Direct exposure to sunlight or strong artifi cial light with high UV components should be
avoided.
Validity Period:
2 years

PRODUCT INFORMATION

For further information or assistance relating to the BIOSENSORS products, please contact:
Legal Manufacturer:
BIOSENSORS INTERNATIONAL PTE LTD
Add: 36 Jalan Tukang, Singapore 619266
Tel: +65-6213-5777
Fax: +65-6213-5737
Email: qara_global_ccp@biosensors.com
RMFC-0006-006 REV. A1
FIG.1
Bladder
FIG.2
FIG.4
EU Representative:
Shanghai International Holding Corp. GmbH (Europe)
Add: Eiffestrasse 80, 20537 Hamburg, Germany
Tel: +49-40-2513175
Fax: +49-40-255726
Email: shholding@hotmail.com
2
0123

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Fc3nFc2nwFc3nw

Tabla de contenido