Descargar Imprimir esta página

Automatic Cleaning/Disinfection With Manual Pre-Cleaning; Manual Pre-Cleaning With A Brush - Aesculap ELAN 4 Instrucciones De Uso

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 40
en
7.8

Automatic cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be serviced
and checked at regular intervals.
7.8.1

Manual pre-cleaning with a brush

Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
Move flexible components (such as sleeves) during cleaning.
Thoroughly clean the product under running water.
Phase II
Move flexible components (such as sleeves) during cleaning.
Brush cannulation with cleaning brush TA011944 and difficult to
access surfaces with a suitable plastic cleaning brush for at least 1 min.
After manual pre-cleaning, check visible surfaces for residue and
repeat the pre-cleaning process as required.
10
T
t
Conc.
Water
[°C/°F]
[min]
[%]
quality
RT (cold) -
-
D–W
RT (cold) -
-
D–W
Chemistry/Note
Until visually clean
Until visually clean

Publicidad

loading

Este manual también es adecuado para:

Ga844