7.8.2

Mechanical alkaline cleaning and thermal disinfection

Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D-W:
Drinking water
FD–W:
Fully desalinated water (demineralized, microbiological, at least of drinking water quality)
*Recommended: BBraun Helimatic Cleaner alkaline
Insert product 1 in the correct position in the Aesculap ECCOS holder
systems, see Fig. B.
Attach the rinsing adapter 14 (GA344211) to the product 1 in the
mount 11.
Then connect the interior rinse device and connect to the rinse connec-
tor of the cleaning/disinfection device/rinse cart.
Attach the Kirschner wire protection sleeve 12 to a rinse hose.
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
7.9

Inspection, maintenance and checks

Allow the product to cool down to room temperature.
Spray product after every cleaning and disinfection with Aesculap
STERILIT Power Systems oil spray GB600 using the oil spray adapter 15
GB600880 (green) for approx. 2 s, see Fig. A.
Note
Aesculap additionally recommends the occasional spraying of moving parts
(such as knobs, couplings) with the Aesculap STERILIT Power Systems oil
spray.
Check the product after each cleaning and disinfection for the follow-
ing: cleanliness, damage, function, abnormal operation noise, excessive
heat or heavy vibration.
Set aside the product if it is damaged.

7.10 Packaging

Follow the instructions for use of the respective packaging and holders
(e.g. instructions for use TA009721 for Aesculap ECCOS holder sys-
tems).
Insert products in the correct position into the Aesculap ECCOS holder
system, see Fig. B.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemistry/Note
-
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
0.5 % working solution
– pH ~ 11*
-
-
at least 10min at max. 120°C

7.11 Steam sterilization

Note
Remove all attached components from the product (tools, accessories)
before sterilization.
Make certain that all external and internal surfaces of the product will
be exposed to the sterilizing agent.
Use a validated sterilization method:
– Steam sterilization using fractional vacuum process
– Steam
sterilizer
DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding
time 5 min
When sterilizing multiple products in one steam sterilizer:
Ensure that the maximum permitted load specified by the manufac-
turer for the steam sterilizer is not exceeded.

7.12 Storage

Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
8.

Maintenance

To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.

Troubleshooting list

Note
For additional information, see the operating instructions for the
ELAN 4 electro control unit GA800 (TA014401).