7.8.2
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D-W:
Drinking water
FD–W:
Fully desalinated water (demineralized, microbiological, at least of drinking water quality)
*Recommended: BBraun Helimatic Cleaner alkaline
►
Insert product 1 in the correct position in the Aesculap ECCOS holder
systems, see Fig. B.
►
Attach the rinsing adapter 14 (GA344211) to the product 1 in the
mount 11.
►
Then connect the interior rinse device and connect to the rinse connec-
tor of the cleaning/disinfection device/rinse cart.
►
Attach the Kirschner wire protection sleeve 12 to a rinse hose.
►
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
7.9
Inspection, maintenance and checks
►
Allow the product to cool down to room temperature.
►
Spray product after every cleaning and disinfection with Aesculap
STERILIT Power Systems oil spray GB600 using the oil spray adapter 15
GB600880 (green) for approx. 2 s, see Fig. A.
Note
Aesculap additionally recommends the occasional spraying of moving parts
(such as knobs, couplings) with the Aesculap STERILIT Power Systems oil
spray.
►
Check the product after each cleaning and disinfection for the follow-
ing: cleanliness, damage, function, abnormal operation noise, excessive
heat or heavy vibration.
►
Set aside the product if it is damaged.
7.10 Packaging
►
Follow the instructions for use of the respective packaging and holders
(e.g. instructions for use TA009721 for Aesculap ECCOS holder sys-
tems).
►
Insert products in the correct position into the Aesculap ECCOS holder
system, see Fig. B.
►
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
►
Ensure that the packaging will prevent a recontamination of the prod-
uct.
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemistry/Note
-
■
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
0.5 % working solution
– pH ~ 11*
-
-
at least 10min at max. 120°C
7.11 Steam sterilization
Note
Remove all attached components from the product (tools, accessories)
before sterilization.
►
Make certain that all external and internal surfaces of the product will
be exposed to the sterilizing agent.
►
Use a validated sterilization method:
– Steam sterilization using fractional vacuum process
– Steam
sterilizer
DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding
time 5 min
When sterilizing multiple products in one steam sterilizer:
►
Ensure that the maximum permitted load specified by the manufac-
turer for the steam sterilizer is not exceeded.
7.12 Storage
►
Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
8.
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting list
Note
For additional information, see the operating instructions for the
ELAN 4 electro control unit GA800 (TA014401).
DIN EN 285
and
validated
en
pursuant
to
11