Guidance and manufacturer's declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
Dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
44
IEC 60601-1-2
test level
+/- 8 kV contact
+/- 2, 4, 8, 15 kV air
+/- 2 kV for power
supply lines
+/- 1 kV for input /
output lines
+/- 1 kV line(s)
to line(s)
+/- 2 kV line(s)
to earth
< 5 % U
(95 % dip
T
in U
) for 0.5 cycle
T
40 % U
(60 % dip
T
in U
) for 5 cycles
T
70 % U
(30 % dip
T
in U
) for 25 cycles
T
< 5 % U
(95 % dip
T
in U
) for 5 sec.
T
30 A/m
Compliance
Electromagnetic
Level
Environment Guidance
+/- 8 kV
If floors are covered with
contact
synthetic material the
+/- 2, 4, 8,
relative humidity shall be
15 kV air
least 30 %.
+/- 2 kV power
Mains power quality shall
supply lines
be that of a typical
N/A
commercial or hospital
environment.
Mains power quality shall
be that of a typical
commercial or hospital
environment.
100 %
Mains power quality shall
reduction
be that of a typical
0.5 period
commercial or hospital
environment.
40 % reduction
for 5 periods
30 % reduction
for 25 periods
100 %
reduction
for 5 sec.
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.