Compatible external equipment
•
External medical grade monitors (video output)
•
External medical imaging recorders (video output and trigger output)
•
USB flash drives
•
Medical USB printer
•
USB powered audio devices that comply with IEC 60601-1, IEC 60950-1 or IEC 62368-1
Note: Verified compatibility with Sony UP-DR80MD digital colour printer for medical
applications. For specifications of connections to external equipment, refer to chapter 12.
Note: IEC 60950-1 and IEC 62368-1 are consumer electronic standards and do not cover patient
safety. Therefore do not touch the accessories while touching the patient and place the
equipment out of reach of the patient.
2.3. Endoscope activation
New endoscope types that are not found on the displaying unit's list of supported endoscopes
(see section 2.2) must be activated with an activation code before they can be used with the
displaying unit. The activation code is entered only once for each endoscope type, and once
an endoscope type has been activated, it can be found on the list of supported visualization
devices. The activation codes are found on Ambu's website via the URL shown on the
displaying unit's screen next to the input field where the activation code is to be entered.
Activate a new endoscope type:
•
Press the Settings tab, then press About.
•
Scroll to Activation codes
for the activation codes.
•
Enter the URL in the address field of the internet browser on your connected device, e.g.
computer, tablet or mobile phone or scan the QR code with your mobile phone.
•
Find the activation code for the endoscope to be activated and enter the code into the
input field below Activation codes
5
and press the question mark
7
.
5
6
6
to find the URL or QR code
7
9