5.3
Inserting the battery into the hand piece
The Translux 2Wave is supplied with a powerful Lithium-ion rechargeable battery.
The battery is supplied separately in the device packaging and must be inserted into the hand piece before initial
operation.
WARNING
Carefully follow all the safety precautions described in the paragraph 3.2.3 (BATTERY – Safety precautions).
CAUTION
NEVER place the hand piece in the charging unit without the battery inserted in the hand piece.
Insert the battery into the hand piece in the sequence described below:
1 Remove the battery from its packaging.
WARNING
Never short circuit the terminals of the battery with any metal object as it could cause
a risk of burns, fire and explosion.
2 Use a flat tool (for example a coin) to unscrew counter-clockwise the battery cap of the
hand piece (Fig. 1 and 2 ).
GB
3 An arrow is marked on the battery case (Fig. 3 ).
NOTE
- 14 -
This arrow indicates the orientation with which the battery must be inserted into the hand piece.
4 Slowly slide the battery with the arrow oriented toward the housing into the hand piece until it stops (Fig. 4 ).
NOTE
If the battery is inserted in the wrong way (i.e. in the opposite direction of the arrow),
the hand piece will not function. In case the battery is inserted in the wrong way,
no damage is caused to the hand piece's electronics.
WARNING
Use original batteries only. Kulzer GmbH will provide only original batteries supplied by Mectron S.p.A. .
The use of other manufacturer batteries or non-rechargeable batteries / primary batteries is a potential
hazard and can cause irreversible damage to the unit.
5 Use a flat tool (for example a coin) to screw clockwise the battery cap of the hand piece.
NOTE
The unit will not function properly if the cap of the hand piece is not screwed on completely.
5.4
Inserting the fibre optic into the hand piece
• Rotate the light guide slightly ( 1 ) and insert it into the hand piece ( 2 ) until final positioning
is reached. The final position is reached when you hear a " click ".
• Mount the protective light cone ( 3 ) on the top end of the light guide.
Positioning the light guide
• Rotate the light guide into the desired position for the polymerisation.
• To make full use of the light intensity provided, place the light guide as close to the composite as possible.
Avoid direct contact with the composite material!
CAUTION – Keep the light guide clean at all times to obtain full light intensity.
The light intensity is considerably reduced if:
– The light guide is not in perfect condition (scratched or chipped).
– The light guide is not correctly placed in the hand piece.
– The light guide has composite stains.
WARNING
A damaged light guide reduces the light intensity and must be replaced with a new one immediately.
Sharp edges may cause serious injury. Use original light guides only. Kulzer GmbH will provide only original
light guides supplied by Mectron S.p.A. .
WARNING – Infection control:
To make treatment as safe as possible for patients and operators, the light guide and the protective light cone
must be cleaned, disinfected and sterilised before each treatment. Follow the instructions listed in paragraph 8
(Cleaning, disinfection and sterilisation) step by step.