Measurement Of The Light Intensity - Kulzer Translux Wave Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 99
Exposure 10 seconds polymerisation
Briefly press the " STAND-BY " button
The 10 second LED is on showing that this program is running.
Once the curing cycle is complete, the unit will automatically turn the light off.
Acoustical signal:
The following acoustical signals (beep) are emitted:
1 beep while press the " STAND-BY " button.
1 beep at the end of the exposure time.
Exposure 20 seconds "Soft-Start" polymerisation
Press and keep pressed for at least one second the " STAND-BY " button
The 20 second LED is on showing that this program is running.
Once the curing cycle is complete, the unit will automatically turn the light off and the unit returns to the 10 second
mode (10 second LED on).
Acoustical signal:
GB
The following acoustical signals (beep) are emitted:
1 beep while press the " STAND-BY " button.
- 14 -
1 beep after 1 second for acoustic indication of " Soft-Start "-mode activation.
1 beep after 10 seconds of exposure time.
1 beep at the end of the exposure time.
NOTE
The exposure cycle can be interrupted at any time, before the exposure time is over,
by pressing again the " STAND-BY " button.
Acoustical signal:
An acoustical signal ( beep ) is emitted, when the " STAND-BY " button is pressed to interrupt the cycle.
6.2

Measurement of the Light Intensity

The charging unit of the Translux Wave is equipped with an integrated light intensity meter ( 8 ).
CAUTION
Before each treatment always check the light intensity by using the light intensity meter
integrated in the Translux Wave charging unit.
• Without applying pressure, place the light guide flat and directly on the testing area ( 8 ).
Activate the unit by pressing the " STAND-BY " button
The light intensity measuring is immediately indicated by the 2-colour LED (Symbol
of the testing area.
GREEN LED : The light intensity is sufficient for the treatment.
YELLOW LED : The light intensity is insufficient and patient/s must not be treated.
CAUTION
If the light intensity is insuffi cient, please carry out the following controls and measures:
– The light guide has not been inserted into the hand piece (see paragraph 5.4 Inserting the fi bre optic
into the hand piece).
– Check the light guide for contamination or defects.
– Clean the light guide of any contamination (see paragraph 8 Cleaning, disinfection and sterilisation)
or replace the damaged light guide with a new one.
Use original Kulzer light guides only.
WARNING
If the measures described above do not result in any improvement, do not use the device for any treatment.
Switch off the unit (disconnect the charging unit from the electrical power source) and ensure that
unauthorised unit re-activation is not possible. Contact a Kulzer's service partner, or technically trained
personnel authorised by Kulzer (see paragraph 14 Service).