Guidance and manufacturer's declaration – Electromagnetic immunity
Signal input / output parts port
The Translux Wave device is designed to work in the electromagnetic environment specifi ed below.
The client or user of the Translux Wave device should always make sure that it is used only in such environment.
Phenomenon
Basic EMC standard
or test method
Electrostatic
IEC 61000-4-2
e)
discharge (ESD)
Electrical fast
IEC 61000-4-4
b) f)
transients / bursts
Surges
IEC 61000-4-5
a)
Line-to-ground
Conducted distur-
IEC 61000-4-6
bances induce by
b) d) g)
RF fi elds
a) This test applies only to output lines intended to connect directly to outdoor cables.
b) SIP/SOPS whose maximum cable length is less than 3 m in length are excluded.
c) Testing may be performed at other modulation frequencies identifi ed by the RISK MANAGEMENT PROCESS.
d) Calibration for current injection clamps shall be performed in a 150 Ω system.
e) Connectors shall be tested per 8.3.2 and Table 4 of IEC 61000-4-2:2008. For insulated connector shells, perform air discharge testing to the connector
shell and the pins using the rounded tip fi nger of the ESD generator, with the exception that the only connector pins that are tested are those that can be
contacted or touched, under conditions of INTENDED USE, by the standard test fi nger shown in Figure 6 of the general standard, applied in a bent or straight
position.
f) Capacitive coupling shall be used.
g) If the frequency stepping skips over an ISM or amateur radio band, as applicable, an additional test frequency shall be used in the ISM or amateur radio
band. This applies to each ISM and amateur radio band within the specifi ed frequency range.
h) r.m.s., before modulation is applied.
i) The ISM (industrial, scientifi c and medical) bands between 150 kHz and 80 MHz are 6, 765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to
27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0, 15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz
to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz,
28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Immunity test levels
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±1 kV contact
100 KHz repetition frequency
± 2kV
3 V
h)
0,15 MHz - 80 MHz
h)
6 V
in ISM bands between
i)
0,15 MHz and 80 MHz
c)
80 % AM at 1 KHz
Electromagnetic environment – Guidance
Floors must be wood, concrete or ceramic.
If the fl oor is lined with synthetic materials,
the relative humidity should be at least
30 %
The quality of the power supply should be
that of a typical commercial or hospital
environment.
The quality of the power supply should be
that of a typical commercial or hospital
environment.
Portable or mobile RF communication
device must not be used in close proximity
of the product, including its cables, except
when these respect the distances of sepa-
ration recommended and calculated from
the equation applicable to the frequency of
the transmitter.
GB