Test specifi cations for enclosure port immunity to RF wireless communications equipment
The Translux Wave device is designed to work in an electromagnetic environment in which RF radiated disturbances are
kept under control. The purchaser or user of the Translux Wave device can contribute to prevent electromagnetic inter-
ference by ensuring a minimum distance between the RF mobile and cordless communication device (transmitters) and
the Translux Wave device, as recommended below, in relation to the maximum output power of the radio-communication
devices.
Test frequency
Band
(MHz)
(MHz)
385
380 – 390
450
430 – 470
710
745
704 – 787
780
GB
810
870
800 – 960
- 28 -
930
1720
1845
1700 – 1990
1970
2450
2400 – 2570
5420
5500
5100 – 5800
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
Translux Wave may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the Translux Wave device, including cables
specifi ed by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
a)
Service
a)
Modulation
Pulse modulation
TETRA 400
18 Hz
FM
GMRS 460,
± 5 kHz deviation
FRS 460
1 kHz sine
LTE Band 13,
Pulse modulation
17
217 Hz
GSM 800/900,
TETRA 800,
Pulse modulation
iDEN 820,
18 Hz
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
Pulse modulation
DECT; LTE
217 Hz
Band 1, 3, 4,
25; UMTS
Bluetooth,
WLAN, 802.11
Pulse modulation
b/g/n, RFID
217 Hz
2450, LTE
Band 7
WLAN 802.11
Pulse modulation
a/n
217 Hz
b)
Maximum
power (W)
b)
1,8
c)
2
b)
0,2
b)
2
b)
2
b)
2
b)
0,2
Distance (m)
Immunity test
level (V/m)
0,3
27
0,3
28
0,3
9
0,3
28
0,3
28
0,3
28
0,3
9