5.0 REAR PANEL INTERFACES

The rear panel has connectors for the data interface, fuse holder, security
lock location and power supply input.
5.1

POWER

Make certain your PMB is compatible with the mains power supply in your
area. If the voltage is not the same the analyzer can be damaged. Contact
your supplier if you require assistance.
The power connection uses a standard IEC C13 socket on the cable with a
plug suitable for the location. Do not use adaptors if the plug does not match
your power supply.
The fuse should only be replaced with a time lag type fuse with high breaking
capacity, 5Amp current rating. The fuse is a standard 5 x 20mm size.
replace the fuse remove all power from the analyzer, unscrew the fuse holder
cover to extract the old fuse, replace with a new fuse. If the fuse should blow
again, the analyzer must be serviced by a qualified service representative.
There are no user serviceable parts inside. Contact your supplier.
5.2

DATA INTERFACES

The PMB has data interfaces for a variety of applications.
The RS232 interface is a general purpose serial interface for communications
with the analyzer, using a DE-9P type connector.
details of the interface.
 