10.1.4
PASSCODES
To enable the security features in the PMB it is necessary to set passcodes.
There are 2 passcodes called Operator Passcode and Supervisor Passcode.
The Operator Passcode allows an authorised user to operate the basic
weighing functions of the analyzer but will not allow access to the Supervisor
Menus if the Supervisor Passcode has been set.
To change or disable a Passcode it will be necessary to enter the current
passcode.
Press [Setup/Enter] when "PASSCODES" is displayed to enter this section.
OPERATOR
SUPERVISOR
10.1.4.1
FORGOTTEN PASSCODES
Keep a record of the passcode to ensure you can access this section again.
If however you have forgotten your passcode you can still gain access by
entering a universal code.
If you have forgotten the current passcode a code of "15" will always allow
you to enter the Supervisor area.
Using the Supervisor menus go to the PASSCODE section and reset the
operator or Supervisor passcode using the "15" code as the old number when
requested.
52
|
P a g e
Enter the current passcode (OLD) first then enter a new
passcode if desired. A passcode set to zero will disable
the security feature and allow unlimited access.
First enter the current passcode (OLD) and then enter a
new passcode if desired. A passcode set to zero will
disable the security feature and allow unlimited access.
© Adam Equipment Company 2017