The configuration file is a text file that would appear as the left column below.
All configuration files will be in English regardless of the language selected
for the analyzers.
Baud Rate=2;
Parity=0;
Print Test=1;
Output Format=0;
Continuous=1;
Interval=20;
Date Format=1;
User ID=1234567890;
Key Beeper=1;
Test Beeper=1;
Backlight=2;
Filter=1;
Stability=2;
Auto Zero=3;
Operator=000000;
Supervisor=000000;
Calibration Report=1;
Language=1;
If there is a problem loading a configuration or PST file from the USB the
display will show the message FAIL.
memory in the reports folder. Open the report file using Notebook or other
word processor to read the message. The message will give a brief
description of the reason for the failed attempt. A line of data may have an
incorrect value or command or it may not be possible to complete the
command as it is restricted to the service personal.
Try correcting the command or removing the line the command is on and
repeating the loading procedure.
 
55
|
P a g e
#0=2400,1=4800,2=9600,3=19200,4=38400,default=1;
#0=none,1=Even,2=Odd
#0=off,1=on, default=1;
#0=complete 1=summary ,default=0;
#0=off,1=on, default=0;
#00-99 ,default=1;
#0=Europe,1=USA,2=ASIA default=0;
#length=10 alpha-numeric characters;
#0=off;1=on default=1;
#0=off;1=on default=1;
#0=off,1=on,2=auto
#0=slow,1=normal,2=fast, default=1;
#0=1d,1=2d,2=5d,3=10d
#0=off,1=1d,2=2d,3=5d
#operator password
#Supervisor password
#0=off,1=on
default=1;
#0=English, 1=German, 2=French, 3= Spanish
A report will be written on the USB
© Adam Equipment Company 2017
,default= 0;