"Time" selected
Enter the time the test should run. The test will stop when this time is
reached.
Display will show "Min XX". This sets the minutes to run a test. The range is
04 to 99 minutes.
"Stable" selected
Enter the conditions to be used to determine if the results are stable. The
amount of weight change to be used to determine if the test is finished,
followed by the time to be used for the weight to change. For example 0.006g
in 15 seconds would be a typical setting. Longer times or smaller weight
changes could be used to ensure better stability or shorter times or larger
weight changes to ensure a faster determination. For example 0.006g in 5
seconds will end a test quicker (but possibly less accurately) than 0.002g in
15 seconds. The user sets the weight first, then the time.
Display shows "STAB RANGE" "
change. Set from 0.001 to 0.099 for PMB 53 or 0.01 to 0.99 for PMB 202.
Then the display shows "Stop Time" " Sec XX". Range is 01 to 99 seconds.
STABLE
St0p
Time/Stable will combine these 2 settings.
TIME/ STAB
St0p
24
|
P a g e
STOP TIME
Min 30
XXX". Set the range for the weight
STAB RANGE
02
STOP TIME
min 30
STAB TIME
sec15
STOP RANGE
02
© Adam Equipment Company 2017
STOP TIME
sec15