6.1.1 MANUALLY SETTING TEST PROCEDURES
Press the [Test/] key to set the test procedure. During the process use the
numeric entry procedure described in section 4.1 to enter or change a setting.
Pressing [Print/] at any time during this procedure will print a list of the
current settings. See 6.1.2
Pressing [0/T / Esc] will always escape back to normal weighing without
a change to the parameter being displayed. Any parameter changed before
this time will be accepted.
PROCEDURE
Press [Test/] to show "rESULt" with the current setting displayed,
Options are "Moisture", "Solid", "ATRO Moist", "ATRO Solid"
Use arrow keys, [Dspl/] and [Test/], to change and press [Setup/Enter]
to continue.
MOISTURE
result
Heating Control, show "HEAt"
Options are: "Single", "Ramp", "Step"
The next settings will depend upon which heating option was selected:
"Single" Selected
Show "Single" "123 C" with the left digit flashing. The user can use the
[PST/] key to select another digit or the [Dspl/] and [Test/] to set the
temperature, Range is 050 C to 160 C
21
|
P a g e
ATRO MOIST
result
SINGLE
Keat