11.5 ANALYZER CONFIGURATION THROUGH THE SERIAL
INTERFACE
The analyzer can use the serial interface to set parameters for weighing test
and for analyzer configuration. The parameters are set by sending a line of
data that has the description and the setting desired.
It is possible to create a text file that contains all the data necessary to set he
test conditions for a particular test (or analyzer configuration) and send this to
the analyzer.
PARITY=0 are both accepted.
11.5.1
TEST PARAMETERS
The parameters that control a test are listed below.
parameter it is necessary to send a command for the parameter to modify
along with the value to change.
For example to change the test parameters to use a single heat and set the
temperature to 115ºC send 2 commands.
Heat=0<cr><lf>
Single Temp=115<cr><lf>
All commands are in English regardless of the language selected for the
analyzer operation. The commands can be sent in both upper case letters
and lower case letter, i.e. "!KT" or "!kt" are both acceptable.
Moisture Result=1;
Heat=1;
Single Temp=100;
Step temp1=050;
Stem Time1=30:00;
Step temp2=080;
Stem Time2=30:00;
Step temp3=050;
Stem Time3=30:00;
Ramp Temp=100;
Ramp Time=30:00;
Rapid=0;
Interval Time=20;
Stop mode=0;
Stop time=10;
Stable Range=0.010;
Stable Time=10;
Start mode=0;
savepos=01;
60
|
P a g e
The commands are not case sensitive, i.e. parity=0 and
#0=%M, 1=%S, 2=%Atro_M, 3=%Atro_S
#0=Single, 1=Ramp, 2=Step
#050C to 160C,
#050C to 160C,
#Time 03:00 to 99:00
#050C to 160C, 000C is off
#Time 03:00 to 99:00
#050C to 160C, 000C is off
#Time 03:00 to 99:00
#050C to 160C,
#Time 03:00 to 99:00
#Rapid heat 0=Off, 1=On, 2=Manual
#05 sec to 99 sec
#0=Time,
#time = 03 to 99 min
#grams default=0.010 for PMB 53, 0.10 for PMB 202;
#seconds, 01-99
#0=Manual
#memory location 01 to 49;
1=Stable, 2=Stab+Time
1=Auto
© Adam Equipment Company 2017
To change a test
default=0;
default=0;
default=100;
default=100;
default=30:00;
default=100;
default=30:00;
default=100;
default=30:00;
default=100;
default=30:00;
default=0;
default=15;
default=3;
default=20;
default=02;
default=0;