5.7
Connecting one or more devices with CAN bus communication
-
Connect up the line connecting module and driver housing. Then hand-tighten all
cap nuts.
-
Set a unique address for each device at the 'CONFIG' port using DIP switches 1 – 6
(see chapter 7.7)
-
The last device in the chain must be activated as a terminator by setting DIP switch
8 to the 'ON' value
-
Connect the first device to the communication connecting line.
-
All other devices can be connected directly from the CAN Out connection to the
next 'Dig I/O' input using the 'CAN connecting line' available as an option (as a
'daisy chain')
-
The DC connecting line must be connected separately to each device.
Sample illustration of a single device with CANopen communication (left) and a row of
three devices with CANopen communication (right), each with the necessary settings at
the Config port. The removable module carriers, module connecting lines and gas lines
are not shown here for the sake of clarity.