Jotec E-vita OPEN NEO Instrucciones De Uso página 26

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 72
EN
Distal aortic diameter [mm]
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
Table 3: Recommended oversizing guideline for aneurysm.
Aortic diameter [mm]
In case of acute dissection, it is recommended to choose no oversizing.
Table 4: Recommended oversizing guideline for dissection.
In case of distal extension, the distal end of the E-vita OPEN NEO stent graft part represents an
adequate landing zone. The overlapping should be at least 45 mm. The overlap marker indicates
the required overlapping.
26
Aneurysm
Recommended diameter
stent graft part [mm]
Dissection
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
26
26
28
28
30
30
33
33
33
36
36
36/40
40
40
40
Recommended diameter stent graft part [mm]
Minimum length of sealing
zone
28 mm
33 mm
22
22
24
24
26
26
28
28
30
30
33
33
33
36
36
36
36
40
40
40
40

Publicidad

Tabla de contenido
loading