Battery power
Device can be used on battery power to provide the user greater mobility. A fully charged battery can last up to
20 hours, dependent on pressure setting.
During therapy, if battery level is low, the device can be plugged into AC mains power without interrupting therapy.
Battery will recharge while therapy continues to be delivered.
Battery status indication
RENASYS™ GO contains a lithium ion rechargeable battery. There are several battery level indicators on this device:
Indicator
Display screen when device is
locked for homecare use.
Homecare lock feature is available only
on the RENASYS GO: REF 66801496
Battery level
Battery is fully charged and has up to
20 hours active therapy remaining.
Battery status top indicator will
illuminate solid green.
Battery has up to 10 hours active
therapy remaining.
Battery status top indicator will flash
green.
Battery has up to 6 hours active
therapy remaining.
Battery status top indicator will flash
green and battery status bottom
indicator will illuminate solid yellow.
Battery has up to 3 hours active
therapy remaining.
Battery status bottom indicator will
illuminate solid yellow.
Low battery alarm state (see alarm
section).
Battery has up to 1 hour active
therapy remaining.
Battery status bottom indicator will
illuminate solid yellow.
Low battery alarm state (see alarm
section).
Battery has only 2 minutes of active
therapy remaining.
Battery status bottom indicator will
illuminate solid yellow.
Low battery alarm state (see alarm
section).
17
Action
No action required.
No action required.
No action required.
Audio alarm can be paused
by pressing the Audio
pause button on user
interface.
Plug device into AC mains
power to charge battery.
Device should be plugged
into AC mains power to
charge battery and ensure
continued delivery of
therapy.
Device should be plugged
into the AC mains power or
therapy will discontinue.
EN