smith&nephew RENASYS GO 66801244 Manual De Instrucciones página 21

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 130
EN
Inactive alarm:
Device has been left in standby mode for longer than 15 minutes. Either select pressure setting and start therapy or
turn device off until therapy is required.
Alarm state
Status/alarm indicator:
Inactive
illuminates solid yellow
Inactive
Status/alarm indicator:
audio paused
illuminates solid yellow
Device not calibrated alarm:
Device detects it is not been calibrated or has failed to calibrate. Device will not operate.
Alarm state
Device not
Status/alarm indicator:
calibrated
illuminates solid yellow
Device failed alarm:
Device has an unrecoverable error. Device will not operate. Contact your Smith & Nephew authorized representative.
Alarm state
Unrecoverable
Status/alarm indicator:
error
illuminates solid yellow
Audio pause
If an alarm state is present, the audible alarm may be paused for 2–3 minutes by pressing the audio pause button.
Alarm states that may be paused by pressing the Audio pause button include:
High flow/leak, Low vacuum, Complete blockage/canister over-capacity, Low battery and Inactive
Note: Inactive alarm can only be paused once.
Visual indication
! Attention
! INACTIVE
! AUDIO PAUSED
! INACTIVE
Visual indication
! DEVICE NOT
! CALIBRATED
Visual indication
! DEVICE FAILED
! Please Return
22
Display screen
Beeps 2 times every
20 seconds
Audio may be paused
once
Paused for 2–3 minutes
Display screen
Beeps 2 times every
20 seconds
Display screen
None
Audible alarm
Audible alarm
Audible alarm

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Renasys go 66801496

Tabla de contenido