EN
Power requirements
Device input voltage
Device input power
Mains adapter
Fuses
Power cord
Dimensions
Weight
Battery
Operating time
Type
Charging time
Safety protection
Protection against electric shock
Patient protection
Ingress protection
Environmental
Storage and transport
Operational temperature
Relative humidity
Atmospheric pressure
Compliance
Standards
Compliance
Standards
21 VDC
36 Watts
Smith & Nephew REF 66800161 or 66801558
Input: 100–240VAC, 50/60HZ, 34–58VA
Output: 21VDC, 1.71A, 36W
Internal electronic fuse, not user changeable
Length: 4.5m maximum (Class I)
1.8m maximum (Class II)
7 x 8.3 x 3.5in. / 175 x 210 x 85mm
2.4lbs / 1.1 kg
20 hours (therapy)
Lithium ion
3 hours
Class I and Class II
Type BF
IP22
14 to 131°F / -10 to 55°C
41 to 95°F / 5 to 35°C
30% to 70% RH
700 mbar to 1,060 mbar
CONFORMS TO
AAMI STD ES60601-1
IEC STD 60601-1
IEC STD 60601-1-8
IEC STD 60601-1-11
CERTIFIED TO CAN/CSA STD C22.2 No. 60601-1
IEC STD 60601-1-2