Safety alarms

General
All alarms for RENASYS™ GO are determined to be "Low Priority," indicating that operator awareness is required
(IEC 60601-1:2005, 3rd edition and IEC 60601-1-8:2006).
In an alarm state with the audio paused, if a different alarm state occurs, the audio pause will be canceled, the
audio will sound and the display screen will indicate the new alarm state. When multiple alarm states are present,
the device's display screen will alternate between the display information for each alarm activated.
Cautions – Lack of alarms:
RENASYS GO device is equipped with high flow/leak, low vacuum and complete blockage/canister over-capacity
alarms to alert users to these events occurring during therapy. These alarms are designed to activate based on
changes in pressure status detected by device. However, there are scenarios that may occur during therapy that
can impact alarm functionality. Therefore, it is important that patient, device and wound dressing are monitored at an
appropriate interval to ensure therapy is being delivered.
Over vacuum alarm:
The system has encountered an excessively high vacuum (of >235mmHg) and device will stop delivering therapy.
Event must be investigated by healthcare provider immediately and corrected.
Alarm state
Status/alarm indicator:
Over vacuum
illuminates solid yellow
High flow/leak alarm:
Device detects a significant leak in the system. Once corrected, alarm will automatically reset and
status/alarm indicator will return to solid green.
Alarm state
Status/alarm indicator:
High flow/leak
illuminates solid yellow
High flow/leak
Status/alarm indicator:
Audio paused
illuminates solid yellow
Low vacuum alarm:
Vacuum level is lower than set point of NPWT by >15mmHg. Once corrected, alarm will automatically reset and
status/alarm indicator will return to solid green.
Alarm state
Status/alarm indicator:
Low vacuum
illuminates solid yellow
Low vacuum
Status/alarm indicator:
Audio paused
illuminates solid yellow
Visual indication
! THERAPY STOP
! OVER VACUUM
Visual indication
! WARNING
! LEAK
! AUDIO PAUSED
! LEAK
Visual indication
! WARNING
! LOW VACUUM
! AUDIO PAUSED
! LOW VACUUM
19
Display screen
Beeps 2 times every
20 seconds
Audio cannot be paused
Display screen
Beeps 2 times every
20 seconds
Audio may be paused
Paused for 2–3 minutes
Display screen
Beeps 2 times every
20 seconds
Audio may be paused
Paused for 2–3 minutes
EN
Audible alarm
Audible alarm
Audible alarm