smith&nephew RENASYS GO 66801244 Manual De Instrucciones página 20

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 130
Low battery alarm:
The system is operating on battery power and has detected depletion of available battery energy. If the power
adaptor is plugged into AC mains power and connected to the device, the alarm will automatically reset.
Alarm state
1. Status/alarm indicator:
illuminates solid yellow
Low battery
2. Battery status bottom
indicator: illuminates solid
yellow
1. Status/alarm indicator:
illuminates solid yellow
Low battery
2. Battery status bottom
(Very low)
indicator: illuminates solid
yellow
1. Status/alarm indicator:
illuminates solid yellow
Low battery
2. Battery status bottom
(Extremely low)
indicator: illuminates solid
yellow
1. Status/alarm indicator:
illuminates solid yellow
Low battery
2. Battery status bottom
audio paused
indicator: illuminates solid
yellow
Battery failed alarm:
The system has detected a failure of the battery in the device. Therapy can only be continued by keeping the device
plugged into AC mains power.
Alarm state
1. Status/alarm indicator:
illuminates solid yellow
Battery failed
2. Battery status bottom
indicator: illuminates solid
yellow
Visual indication
! Attention
! Low Battery
! RECHARGE
! V. Low Battery
! RECHARGE
! Ex. Low Battery
! AUDIO PAUSED
! Low Battery
Visual indication
! BATTERY FAILED
! BATTERY FAILED
(shown when locked
for homecare use)
21
Display screen
Display screen
l100 mmHg
l100 mmHg ***
EN
Audible alarm
Beeps 2 times every
20 seconds
Audio may be paused
Beeps 2 times every
20 seconds
Audio may be paused
Beeps 2 times every
20 seconds
Audio may be paused
Paused for 2–3 minutes
Audible alarm
None

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Renasys go 66801496

Tabla de contenido