Electromagnetic Compatibility
EN 60601-1-2 Classification Information
• Internally Powered Equipment
• Continuous Operation
• Ordinary Equipment
• Class II
Guidance and Manufacturer's Declaration - electromagnetic emissions
The Precision Novi System intended for use in the electromagnetic environment specified
below. The customer or user of the IPG should assure that it is used in such an environment.
Emissions Test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
Flicker emissions
IEC 61000-3-3
Compliance
Group 1
The Precision Novi System uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Class B
The Precision Novi System is suitable for
use in all establishments. Including domestic
Class B
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
Complies
purposes.
Precision Novi™ System Information for Prescribers
Electromagnetic Compatibility
Electromagnetic Environment Guide
90962628-03