Automatic cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be serviced
and checked at regular intervals.
7.7.1

Manual pre-cleaning with a brush

Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
For Kirschner wire attachment GB641R: Set the adjustment sleeve 20
to the greatest Kirschner wire diameter.
Phase I
Mobilize non-rigid components during cleaning.
Thoroughly clean the product under running water.
Phase II
Mobilize non-rigid components during cleaning.
Brush cannulation with cleaning brush TA011944 and difficult to
access surfaces with a suitable cleaning brush made of plastic for at
least 1 min.
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available on Aesculap Extranet at
https://extranet.bbraun.com).
10
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
-
-
RT (cold)
-
-
Water
Chemistry/Note
quality
D–W
Until visually clean
D–W
Until visually clean