Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Product becomes too hot
Insufficient power
Loud running noise
Attachment does not
attach or detach.
Tool does not attach or
detach.
For GB641R: Kirschner
wire does not insert into
the Kirschner wire chuck
For GB641R: Kirschner
wire not turning
12
Cause
Excessive use
Defective attachment drive/ball
bearings
Fall damage, product defective
Blunt tool
Product is operating in left rota-
tion mode
Attachment defective
Blunt tool
Defective attachment drive/ball
bearings
Attachment not compatible
Connection on attachment
deformed/defective
Coupling on drill deformed/defec-
tive
Sleeve on drill hard to move
Incompatible tool
Tool connector deformed/defec-
tive
Coupling on attachment
deformed/defective
Tool coupling dirty
Kirschner wire chuck not set cor-
rectly
Kirschner wire chuck not set cor-
rectly
Open the clamping lever
Clamping lever not pulled back
with sufficient force
Detection
Attachment very hot
Attachment very hot
Attachment very hot
Heating of tool and product
Serrated tool run in counterclock-
wise mode
Attachment very hot
Tool blades worn out
Loud, perceptible noise during
operation
Attachment doesn't engage
Attachment is difficult to couple
or cannot be coupled
Attachment is difficult to couple
or cannot be coupled
Attachment doesn't engage
Tool does not click into place
Tool is difficult to couple or can-
not be coupled
Tool is difficult to couple or can-
not be coupled
Tool is difficult to couple or can-
not be coupled
Diameter of the Kirschner wire is
not consistent with the scale set-
ting
Diameter of the Kirschner wire is
not consistent with the scale set-
ting
Open the clamping lever
Kirschner wire not turning
Remedy
Follow operating instructions (operating mode).
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every
sterilization.
Have the manufacturer repair the attachment.
Have the attachment repaired by the manufacturer
Change tool.
Use serrated tools only in clockwise mode
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every
sterilization.
Follow operating instructions (operating mode).
Have the attachment repaired by the manufacturer
Change tool.
Follow operating instructions (preparation, care).
Preventative: lubricate attachment before every
sterilization.
Have the manufacturer repair the attachment.
Use the compatible attachment for drill.
Have the manufacturer repair the attachment.
Have the manufacturer repair the drill.
Turn and hold the drill sleeve, then attach the
attachment. Turn the sleeve back if necessary.
Preventative: lubricate the sleeve on the drill
before each sterilization.
Use the tool compatible with the attachment.
Use a new tool.
Have the manufacturer repair the attachment.
Clean tool or use a new tool.
Clean the attachment.
Set the adjustment sleeve to the right Kirschner
wire diameter.
Set the adjustment sleeve to the right Kirschner
wire diameter.
Close the clamping lever.
Pull back the clamping lever forcefully and hold it.