12.3 Operating mode

Operating mode

12.4 Environmental conditions

Operation
Temperature
10 °C to 27 °C
Relative
30 % to 75 %
humidity
Atmospheric
700 hPa to 1 060 hPa
pressure
Operation with non-periodic load and
speed changes (type S9 pursuant to
IEC EN 60034-1)
Right/left rotation:
60 second application, 60 second
pause
20 repetitions
30 min cooling time
Max. Temperature 48 °C
Milling (clockwise/counterclockwise):
30 second application, 30 second
pause
8 repetitions
30 min cooling time
Max. Temperature 48 °C
Drilling (oscillation):
60 second application, 60 second
pause
4 repetitions
30 min cooling time
Max. Temperature 48 °C
Saw mode with GB660R:
30 second application, 60 second
pause
3 repetitions
30 min cooling time
Max. Temperature 48 °C
Storage and transport
-10 °C to 50 °C
10 % to 90 %
500 hPa to 1 060 hPa

13. Disposal

Note
The user institution is obliged to process the product before its disposal, see
Validated reprocessing procedure.
Adhere to national regulations when disposing of or recy-
cling the product, its components and its packaging!
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
Detailed information concerning the disposal of the product is avail-
able through your national B. Braun/Aesculap agency, see Technical
Service.
15