Issue
Tool not moving
Oil spray adapter not
attachable
10. Technical Service
Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
WARNING
►
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
►
Do not modify the product.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
►
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
11. Accessories/Spare parts
Art. no.
Designation
GB243R
ECCOS set for 2 large machines
GB600
STERILIT Power Systems oil spray
GB600840
Oil spray adapter
TA011944
Cleaning brush
GB496R
ECCOS fixation for 3 attachments
GB497R
ECCOS fixation for one attachment
GA031R
Key for large Jacobs chuck
GA062R
Key for small Jacobs chuck
TA014541
Instructions for use for Acculan 4 attachments
(flyer)
Cause
Attachment not fully attached to
the drill
Tool not fully attached to the
attachment
Attachment defective
Drill defective
The safety catch of the drill is in
the OFF position
Oil spray adapter not compatible
Detection
Attachment can be pulled out of
the drill
Tool can be pulled out of the tool
coupling
Drill turning, but attachment is
not turning.
Drill not turning
Safety catch is in OFF position
Oil spray adapter not attachable
12. Technical data
12.1 Classification acc. to Directive 93/42/EEC
Art. no.
Designation
GB176R
Zimmer chuck with Hudson/Zimmer shaft
GB184R
Harris chuck with AO large shaft
GB620R
Drill attachment keyless Jacobs chuck
GB621R
Drill attachment large Jacobs chuck
GB623R
Drill attachment AO small
GB628R
Adapter AO large to DHS/DCS step drill
GB639R
Drill attachment small Jacobs chuck
GB641R
Kirschner wire attachment
GB645R
Drill attachment for Synthes radiolucent angle
transmission
GB657R
Medullary reamer attachment AO large
GB660R
Sagittal saw attachment
GB663R
Drill attachment Trinkle
GB664R
Drill attachment Aesculap hexagon
GB665R
Drill attachment Hudson/Zimmer
GB667R
Reamer attachment large Jacobs chuck
GB668R
Reamer attachment AO large
GB669R
Reamer attachment Hudson/Zimmer
GB670R
Reamer attachment Harris
XF457R
DIN chuck with AO large shaft
12.2 Performance data, information about standards
Rotational direction
Conforming to stan-
dard
The product was subject to 500 preparation cycles as a test by the manu-
facturer and has passed.
The performance data for the individual attachments can be found in the
table below.
Remedy
Correctly attach the attachment and perform a
function test.
Correctly attach the tool and perform a function
test.
Have the manufacturer repair the attachment.
Have the manufacturer repair the drill.
Turn the safety catch into the ON position.
Use the compatible oil spray adapter for GA344.
Right and left rotation, oscillation
IEC/DIN EN 60601-1
Class
IIa