CONNECTIONS
4-20mA OUTPUT
page 1 of 2
The pump is equipped with a 4-20 mA output signal.
20
18
16
14
12
10
8
6
4
0
25
50
% Pump Speed
+
4-20mA Output
from pump
–
The loop impedance must be less than 300 ohms.
Do not short the 4-20mA output loop. Doing so will cause the pump to shut off and may damage
the pump.
To ensure proper signal output, always calibrate the output signal.
Connection Terminals
• Connect load positive to 4-20mA OUT (+), position #15 on the bottom row.
• Connect load common to 4-20mA OUT (-), position #14 on the bottom row.
1
2
3
4
Output
Output
Standby
Relay
Relay
+
(
1
1
Output
Output
4-20mA
Relay
Relay
OUT
−
3
3
(
11
12
13
14
S Series FW 4.01.02 US and Canada 800.683.2378, International 904.641.1666 www.stenner.com
continued
This signal corresponds proportionally to pump speed with 0% pump
speed equaling 4.0mA and 100% pump speed equaling 20.0mA. The
signal response is not adjustable.
75
100
5
6
7
8
4-20mA
Pulse/
Standby
0-10VDC
IN
PPM/Hall
−
+
)
(
)
(
signal)
+
(
signal)
signal
Leak
Leak
Flow
4-20mA
Detect
OUT
Detect
Detect
+
)
)
(
External
External
15
16
17
18
The pump sources the voltage for the output
signal loop at 24VDC. The pump will control the
+
magnitude of current on the loop (from 4 to
Load
20mA) according to the speed that the pump is
running at. The maximum loop impedance (load
on the output signal loop) is 300 ohms, which
–
includes the load plus any resistance due wire
length, connections, etc.
9
10
Signal
Hall
+
Common
12VDC
Flow
Flow
Detect
Detect
−
+
(
)
12VDC
19
20
74