5.4 Installing the E32A SSR Interface
5.4.1 Mechanical installation
Consult Figure 5-9 for the overall dimensions.
Wire the cable for earthing (see sec. 5.4.2 "Earthing").
Connect the module to the guide and secure it by pressing until the catch snaps into place.
48mm
Figure 5-9. E32A SSR overall dimensions.
5.4.2 Earthing
Earth the equipment by connecting the
dedicated terminal
back, to the machine earth bar. The
connection must be as short as
possible. Use a yellow/green wire with
a cross-section of at least 4mm
5.4.3 E32A SSR wiring
The interface requires two cables, one for connection to the arm and the other for connection to the
machine PLC/CNC. Use the connectors supplied.
The following paragraphs describe the connections to J1 and J2.
116mm
, located on the
2
.
Terminal
M4 nut
Wire terminal
Yellow/green wire
2
4mm
cross-section
Section 5. INSTALLATION