Maximum diameter
of insertion portion
Working length
Working channel
Minimum working
channel width
2
Storage and transportation
Transportation
temperature
Storage temperature
Relative humidity
Atmospheric pressure
1.
Please be aware that the bending angle can be affected if the insertion cord is not kept straight;
or endoscopic instruments have been inserted.
2.
There is no guarantee that accessories selected solely using this minimum working channel width will be
compatible in combination.
3.
Storage under higher temperatures may impact shelf life.
5. Troubleshooting
If problems occur with the system or product, please use this troubleshooting guide to identify
the cause and correct the error.
Problem
No live image on the
Ambu displaying
unit, but the user
interface is present
on the Ambu
displaying unit or
the image shown
is frozen.
Low picture quality.
Absent or reduced
flow of fluid, e.g.
sterile water or saline
solution, or difficulty
inserting the
endoscopic
instrument through
the working channel.
10
Max 18 Fr/
6.0 mm (0.24")
388 mm +12/-8 mm
(15.3" +0.5"/-0.3'')
Min. 6.6 Fr/
2.2 mm (0.086")
-10 – 55 °C
(14 – 131 °F)
10 – 25 °C
3
(50 – 77 °F)
10 – 85 %
50 – 106 kPa
Possible cause
The aScope 5 Cysto HD
is not connected to the
Ambu displaying unit.
The Ambu displaying
unit and the aScope 5
Cysto HD are not
communicating correctly.
The aScope 5 Cysto HD
is damaged.
A recorded image
is shown.
Unwanted fluids etc. on
the distal end.
The working channel
is blocked.
The bending section is
not in neutral position.
Illumination method
Sterilisation
Method of
sterilisation
Operating environment
Temperature
Relative humidity
Atmospheric pressure
Altitude
Recommended action
Connect the aScope 5 Cysto HD
to the grey port on the Ambu
displaying unit.
Restart the Ambu displaying unit
(please refer to the Ambu displaying
unit's Instructions for Use).
Replace the aScope 5 Cysto HD with
a new one.
Return to the live image
(please refer to the Ambu displaying
unit's Instructions for Use).
Gently rub the distal tip against the
mucosa. If the lens cannot be cleaned
this way remove the endoscope and
wipe the lens with sterile gauze.
Clean the working channel using a
cleaning brush or flush the working
channel with sterile water or saline
using a syringe.
Move the bending section into
neutral position.
LED
ETO
10 – 40 °C
(50 – 104 °F)
30 – 85 %
80 – 106 kPa
≤ 2000 m