the Locator Tool, ensuring the black center lines
are centered with the valve mechanism (Figure 10).
Repeat Steps 1 through 6.
CAUTION: Do not move the Locator and Indicator
Tools together to obtain or center a number in
the window as this could result in an inaccurate
indication of the performance setting.
7. Insert the Adjustment Tool into the Locator Tool so
that the arrow is pointing to the current performance
setting. With one hand holding the "petals" of the
Locator Tool, turn the Adjustment Tool with the
other hand until the arrow points to the desired
performance setting (Figure 12). The Adjustment Tool
will provide an audible click and a tactile response as
you turn to each setting.
CAUTION: Failure to accurately position the
Locator Tool could result in an inability to adjust
the performance setting. The Locator Tool must be
precisely aligned with both the valve's direction of
flow and the center of the hard valve mechanism
for an accurate adjustment. Alignment can be more
challenging if tissue thickness is >10 mm above the
valve. In these instances, verify the valve position
and orientation with x-ray or fluoroscopy. Refer
to SECTION D: Troubleshooting and SECTION E:
Confirming the Current Valve Setting.
Note: The Adjustment Tool is equipped with a
mechanical stop between Settings 1 and 8. This is to
prevent inadvertent adjustment between the extremes
of the available settings.
Note: Successful adjustment is accomplished by
starting at the valve's current setting and moving
directly to the desired setting.
8. Remove the Adjustment Tool from the Locator
Tool by lifting it straight upwards a minimum of 3 cm
(1.25 in.) before moving it horizontally away from the
Locator Tool in order to avoid inadvertently changing
the valve setting.
9. Repeat Steps 7 and 8 to confirm successful
adjustment of the performance setting. Always
confirm the desired performance setting of the valve.
10. If the desired performance setting is not achieved,
repeat Steps 7 through 9.
11. It is recommended to record the valve setting in the
patient's record and I.D. wallet card (available from
local Codman sales representative).
12. Disinfect the tool kit components (See Disinfecting
the CODMAN CERTAS Tool Kit).
13. Return all tools to their proper locations in the
storage case to prevent damage.
Note: place the Low Profile Locator Tool under the
elastic webbing in the bottom of the storage case
and fully collapse the Adjustable Height Locator
Tool (See Figure 2B).

SECTION D: Troubleshooting

If valve function is adversely affected by accumulations
of biological matter, it might be possible to dislodge the
material and restore proper function through one of the
following methods: