Trouble Shooting
The coagulation out-
put is disabled by
the Electrosurgical
Generator
Message: "Please
Wait, Cooling down"
Message: "Please
Wait, Warming up"
A fault error mes-
sage is displayed
that is not listed
above
31
Proprietary & Confidential - Page 34 of 391 - Uncontrolled if Printed
For certain Instruments, the Electrosurgical Generator stops delivery of energy
for coagulation after a defined period of time.
Release and then depress the blue Footswitch pedal to activate the
coagulation function
The Electrosurgical Generator is cooling down to an acceptable operating
temperature before starting.
Wait until the Electrosurgical Generator has reached acceptable operating
temperature. Note: maximum ambient temperature during operation is 30 °C
(86 °F).
The Electrosurgical Generator is warming up to an acceptable operating
temperature before allowing start-up of the Electrosurgical Generator.
Wait until the Electrosurgical Generator has reached acceptable operating
temperature before turning it ON. Note: minimum ambient temperature
during operation is 10 °C (50 °F).
The Electrosurgical Generator might not have completed its start-up
checks when the AC supply was interrupted, or there has been significant
electromagnetic interference in close proximity.
Switch the Electrosurgical Generator OFF at the Rear Panel and wait for 10
seconds for internal circuits to discharge. Restart the Electrosurgical Generator
and once the ON / Standby LED on the Front Panel is flashing green, press the
ON / Standby Button to allow completion of system self-checks.
Contact Creo Medical if the error persists and report the error or fault message
and any error code that is also displayed.